Platform code5903662
Platform nameAPEX Profiling Float
DescriptionArgo AUSTRALIA
Ascending Profiling Time6.94
Calib Rt Coefficienta1=1, a0=-0.0003
Calib Rt CommentReal-time salinity adjustment
Calib Rt Date20200615000000
Calib Rt EquationPSAL_ADJUSTED = a1 * PSAL + a0
Calib Rt ParameterPSAL
Cycle Time240
Data CentreCS
Deepest Pressure2000
Deploy PlatformAurora Australis
DirectionA
Inst Reference5357
LABEL ENCTD_PRES
LABEL ENCTD_TEMP
LABEL ENCTD_CNDC
LABEL ENFLOAT
Launch Qc1
ParameterTEMP
ParameterPSAL
ParameterPRES
Parameter Accuracy0.002
Parameter Accuracy0.005
Parameter Accuracy2.4
Parameter Resolution0.001
Parameter Resolution0.001
Parameter Resolution0.1
Parameter Unitsdecibars
Parameter Unitspsu
Parameter Unitsdeg C
Parking Pressure1000
Parking Time224.88
Pi NameSusan Wijffels
Platform MakerWRC
Platform ModelAPEX-SBE
Positioning SystemARGOS
Pr Experiment IdArgo Australia
Pr Launch Datetime12/02/2011 19:31:00
Pr Launch Latitude-48.48
Pr Launch Longitude121.77
Pr Probe Code846
Predeployment Calib Coefficientser# = 5192 temperature coeffs: A0 = 0.0001 A1 = 0.0003 A2 = -0.0000 A3 = 0.0000
Predeployment Calib Coefficientser# = 5192 pressure coeffs: PA0 = 0.65633 PA1 = 0.13984 PA2 = -4.1934e-08 PTCA0 = 33.8953 PTCA1 = 0.33371 PTCA2 = -0.00538 PTCB0 = 25.3481 PTCB1 = -0.000175 PTCB2 = 0 PTHA0 = -79.0852 PTHA1 = 0.04798 PTHA2 = -1.197e-07
Predeployment Calib Coefficientser# = 5192 conductivity coeffs: G = -1.0226 H = 0.1467 I = -0.0004 J = 0.0000 CPCOR = -0.0000 CTCOR = 0.0000 WBOTC = 0.0000
Predeployment Calib Equationf = inst freq * sqrt(1.0 + WBOTC * t) / 1000.0; t = temperature [deg C]; p = pressure [decibars]; delta = CTcor; epsilon = CPcor; Conductivity = (g + hf^2 + if^3 + jf^4)/(1+ delta t + epsilon p) Siemens/meter
Predeployment Calib EquationTemperature ITS-90 = 1/ { a0 + a1[lambda nu (n)] + a2 [lambda nu^2 (n)] + a3 [lambda nu^3 (n)]} - 273.15 (deg C)
Predeployment Calib Equationy=thermistor output; t=PTHA0+PTHA1*y+PTHA2*y^2; x=pressure output-PTCA0+PTCA1*t+PTCA2*t^2; n=x*PTCB0/(PTCB0+PTCB1*t+PTCB2*t^2); pressure (psia)=PA0+PA1*n+PA2*n^2
Project NameArgo Australia
Ptt85408
Repetition Rate1
SensorCTD_TEMP
SensorCTD_PRES
SensorCTD_CNDC
Sensor Accuracy0.005
Sensor Accuracy2.4
Sensor Accuracy0.002
Sensor MakerSBE
Sensor MakerDRUCK
Sensor MakerSBE
Sensor ModelSBE41_V3
Sensor ModelDRUCK
Sensor ModelSBE41_V3
Sensor Resolution-0.0001
Sensor Resolution-1e-05
Sensor Resolution0
Sensor Serial No5192
Sensor Serial No5192
Sensor Serial No3155280
Sensor UnitsSiemens/meter
Sensor Unitsdecibars
Sensor Unitsdeg C
Start Date13/02/2011 18:20:23
Start Date Qc1
Surface Time8.06
Trans Frequency401.65 x 10^6
Trans Repetition46
Trans SystemARGOS
Trans System Id02039
Eulerian networksMyOcean eulerian platforms
Available dataFrom 12/02/2011 to 07/03/2022
Selected data

From ... to ... Refresh

Filters

Download time seriesTo download data, please login here
Download vertical profilesTo download data, please login here
Physical parametersNo valid measure for this period
Waterfall profiles